CLEANING AFTER EACH USE
Following the cleaning instructions after each use will prevent any
remaining medication in the bottle from drying resulting in the device not
nebulizing effectively and will help prevent infections.
WARNING
Wash the nebulizer parts after each use. Dry the parts immediately after washing.
1. Remove the inhalation accessory (mouthpiece, mask, or nosepiece) from
the nebulizer kit.
2. Disconnect the air tubing from the nebulizer kit.
3. Gently twist the inhalation top counterclockwise and lift to separate the
nebulizer kit into two sections.
4. Remove the baffle.
5. Discard remaining medication in the medication cup.
WARNING
Always dispose of any remaining medication in the
medication cup after each use. Use fresh medication
each time you use the device.
6. Rinse all the parts of the nebulizer kit and the mouthpiece or nosepiece
with hot tap water. You may also wash these parts in a dishwasher.
Remove the parts after the wash and rinse cycle is completed.
DO NOT wash the mask in a dishwasher.
CAUTION
Place the nebulizer pieces in a basket recommended for dishwasher use. Place
the basket on the top shelf to avoid any damage to the nebulizer kit.
7. Rinse the mask with hot tap water.
8. Hand dry or air dry in a clean
environment using a soft, clean,
lint-free cloth.
9. Assemble the nebulizer kit and store the
nebulizer kit in a dry, sealed bag.
CAUTION
The neb
ulizer kit should be replaced e
v
ery 6 months.
m
a
x
5
3
3
5
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 15