Omron HBP-1300 User Manual Page 31

  • Download
  • Add to my manuals
  • Print
  • Page
    / 38
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 30
29
EN
Other problems
Disposal
Description
As there is a risk of environmental pollution, follow your applicable national and local legal regulations
regarding disposal or recycling of this equipment and batteries.
The main constituents of each part are listed in the table below. As there is a risk of infection, do not
recycle patient attachments such as cuffs, but dispose of them as instructed by your facility’s
procedures and applicable regulations.
Error
code
Priority Cause Solution
E3 Low
Inflated the cuff to 300 mmHg or more
during inflation in “Auscultation Mode”.
When inflating in “Auscultation Mode”,
release the button when the pressure
reaches the desired value.
Over-inflation occurs
If this occurs during measurement, repeat
measurement.
If this occurs when not performing
measurement, contact OMRON
HEALTHCARE.
E40 Medium The battery is depleted.
Recharge the battery, or replace with a new
battery. (page 14)
E41 Medium Battery failed to charge.
Try charging again. If continues to fail,
replace with a new battery. (page 14)
E42 Medium Battery voltage error
Replace the battery with a new battery. If the
error continues, contact OMRON
HEALTHCARE.
Item Parts Material
Package
Box Cardboard
Cushion Cardboard
Bag PE
Main unit and
accessories
Enclosure ABS, PC , SR
Internal parts General electronic components
Battery pack
Outer tube PVC
Cell batteries Nickel-metal hydride
Internal parts General electronic components
Page view 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 37 38

Comments to this Manuals

No comments